3 To Set the Horizontal System RIGOL
DS6000 User’s Guide 3-5
X-Y Mode
In this mode, the oscilloscope automatically turns on all the four channels (CH1 to
CH4) and the screen is divided into two coordinate areas (XY1 and XY2). In the XY1
area (at the left of the screen), the X axis and Y axis track the voltages of CH1 and
CH2 respectively; in the XY2 area (at the right of the screen), the X axis and Y axis
track the voltages of CH3 and CH4 respectively.
The phase deviation of two signals with the same frequency can be easily measured
via Lissajous method. The figure below shows the measurement schematic diagram
of the phase deviation.
According to
B
A
sin
=θ
or
D
C
(θ is the phase deviation angle between the two
channels and the definitions of
A
,
B
,
C
and
D
are as shown in the figure
above), the phase deviation angle is obtained, that is:
B
A
arcsin±
=θ
or
D
C
arcsin±
If the principal axis of the ellipse is within the quadrant I and quadrant III, the phase
II I
III IV
A
B
C
D
The signal must be
centered horizontally
III IV
II I

Table of Contents

Other manuals for Rigol DS6104

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS6104 and is the answer not in the manual?

Rigol DS6104 Specifications

General IconGeneral
BrandRigol
ModelDS6104
CategoryTest Equipment
LanguageEnglish

Related product manuals