7 Protocol Decoding RIGOL
DS6000 User’s Guide 7-5
RS232 Decoding (Option)
RS232 serial bus consists of the transmitting data line (TX) and the receiving data
line (RX).
DeviceA DeviceB
Tx
Rx
Rx
Tx
The industry standard of RS232 uses "Negative Logic", namely high level is logic "0"
and low level is logic "1".
1
0
0
0 0
1 1
1
1
1
By default, RS232 uses LSB (Least Significant Bit) transmission sequence, namely
the lowest bit of the data is transmitted first. While for MSB (Most Significant Bit),
the highest bit of the data is transmitted first.
Bit0
Bit5
Bit2
Bit3
Bit6
Bit1
Bit7
Bit4
t
Bit7
Bit2
Bit5
Bit4
Bit1
Bit6
Bit0
Bit3
t
Endian (LSB) Endian (MSB)
In RS232, baud rate is used to represent the transmitting rate (namely bits per
second) of the data.