16 Specifications RIGOL
DS6000 User’s Guide 16-1
16 Specifications
All the specifications are guaranteed except the parameters marked with "Typical"
and the oscilloscope needs to operate for more than 30 minutes under the specified
operation temperature.
Sample
Sample Mode Real-time Sample, Equivalent Sample
Real Time Sample Rate
5 GSa/s (single-channel)
2.5 GSa/s (dual-channel)
Equivalent Sample Rate 100 GSa/s
Peak Detect
200 ps (single-channel)
400 ps (dual-channel)
Averaging
After all the channels finish N samples at the same time,
N can be 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024, 2048,
4096 or 8192.
High Resolution
12 bit resolution when ≥5 μs/div @ 5 GSa/s (or ≥10
μs/div @ 2.5 GSa/s).
Memory Depth
single-channel: Auto, 14k pts, 140k pts, 1.4M pts, 14M
pts and 140M pts are available
dual-
channel: Auto, 7k pts, 70k pts, 700k pts, 7M pts and
70M pts are available