RIGOL 13 System Function Setting
13-12 DS6000 User’s Guide
2. Fast
After this type is selected, the oscilloscope can output a fast edge signal which
can be used in the self-calibration of the oscilloscope.
3. GND
After this type is selected, the oscilloscope can output a ground level.
4. PassFail
After this type is selected, the oscilloscope will output a pulse signal when failed
waveforms are detected. Connect this signal to other control systems to
conveniently view the test results.
Reference Clock
This oscilloscope can output the internal 10 MHz sample clock from the [10MHz
In/Out] connector on the rear panel and receive an external 10 MHz clock to
synchronize multiple oscilloscopes.
Press UTIL RefClock to select the desired clock type.
ClockOutput: configure the [10MHz In/Out] connector as an output and
output the internal 10 MHz clock.
ClockInput: configure the [10MHz In/Out] connector as an input and the
oscilloscope will receive an external 10 MHz clock input from this connector.
10MHz In/Out