RIGOL 2 To Set the Vertical System
2-4 DS6000 User’s Guide
Bandwidth Limit
Set the bandwidth to reduce the display noise. For example, the signal under test is a
pulse signal with high frequency oscillation.
When the bandwidth limit is turned off, the high frequency components of the
signal under test can pass the channel.
When the bandwidth limit is turned on and is set to 20 MHz or 250 MHz, the high
frequency components that exceed 20 MHz or 250 MHz of the signal under test
are attenuated.
Press CH1 BW Limit and use to set the status of the bandwidth limit (by
default, it is turned off). When the bandwidth limit (20 MHz or 250 MHz) is turned on,
the character "B" will be displayed in the channel label at the bottom of the screen.
You can also press BW Limit continuously to switch the status of the bandwidth
limit.