9 Pass/Fail Test RIGOL
DS6000 User’s Guide 9-3
To Select Source
Press Source to select the channel (channel enabled from CH1 to CH4) to be tested.
During the test, the oscilloscope will judge whether each frame of the waveform in
the source complies with the current test mask and those waveforms that are beyond
the test mask area (namely the waveforms enter the blue area) are considered as
failed waveforms.
Mask Range
Users can define their desired test masks.
Press Range ïƒ Range to select "Screen" or "Cursor" for test. When "Cursor" is
selected, two gray cursor lines appear on the screen. At this point, press Cursor A
and Cursor B; then, use to adjust the positions of the two cursor lines
respectively to determine the test range. Or, press CursorAB and use to adjust
the positions of cursor A and cursor B at the same time. Note that you can press
down
continuously to switch the current cursor.
Press X Mask and Y Mask respectively and turn , the mask lines appear on the
screen. Press Create to apply the mask currently created. The ranges of the
horizontal and vertical adjustments are from 0.02 div to 4 div and 0.03 div to 4 div
respectively.