RIGOL 10 Waveform Recording
10-4 DS6000 User’s Guide
Record Constant On
In the "Open" mode, the oscilloscope records the waveforms of all the channels
currently turned on in real-time until users press RUN/STOP. The maximum
number of frames can be recorded is limited by the memory depth. When the
number of frames recorded reaches the maximum number of frames can be
recorded, the oscilloscope will continue to record the waveform in real-time and the
waveform data recorded formerly will be overwritten by the waveform data recorded
latterly.
Press UTIL  Record  Mode and use to select "Open". The backlight of the
waveform recording key on the front panel turns on (in red) and flashes. The
oscilloscope records the waveform in real-time according to the current memory
depth until users press RUN/STOP. After the oscilloscope stops recording the
waveform, you can play back or analyze the recorded waveform.
Note: In this mode, the "Acquisition Mode", "Sample Mode", "Memory Depth",
"Pass/Fail Test", "Protocol Decoding" and "Time Base Mode" functions are
disabled; other function are not affected.

Table of Contents

Other manuals for Rigol DS6104

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS6104 and is the answer not in the manual?

Rigol DS6104 Specifications

General IconGeneral
BrandRigol
ModelDS6104
CategoryTest Equipment
LanguageEnglish

Related product manuals