RIGOL 9 Pass/Fail Test
9-4 DS6000 User’s Guide
Test and Output
Before the test, you can use the following method to set the output mode of the test
results.
Press Message to select "ON" or "OFF". When "ON" is selected, the test results will
be displayed at the upper-right corner of the screen.
Press StopOnFail to select "ON" or "OFF".
ON: when failed waveforms are detected, the oscilloscope will stop the test and
enter the "STOP" state. At this point, the results of the test remain the same
with the display on the screen (if display is turned on) and the [Trig
Out/Calibration] (if enabled) connector on the rear panel only outputs one
pulse.
OFF: the oscilloscope will continue with the test even though failed waveforms
are detected. The test results on the screen will update continuously (if the
display is turned on) and the [Trig Out/Calibration] connector (if it is enabled)
on the rear panel will output a pulse each time a failed waveform is detected.
Press Output to select "Fail" or "Fail+ ".
Fail: when failed waveforms are detected, there are display and output but the
beeper does not sound.
Fail+
: when failed waveforms are detected, there are display and output and
the beeper sounds (not related to the on/off state of the sound).
Press AuxOutput to quickly turn "ON" or "OFF" the test result output of the [Trig
Out/Calibration] connector on the rear panel. You can also press UTIL
AuxOutput and select "PassFail" to set this output.