RIGOL 2 To Set the Vertical System
2-6 DS6000 User’s Guide
Table 2-1 Probe Attenuation Coefficient
Menu
Attenuation Coefficient
(signal display amplitude : actual signal amplitude)
0.01X
0.02X
0.05X
0.1X
0.2X
0.5X
1X
2X
5X
10X
20 X
50 X
100 X
200 X
500 X
1000 X
0.01:1
0.02:1
0.05:1
0.1:1
0.2:1
0.5:1
1:1
2:1
5:1
10:1
20:1
50:1
100:1
200:1
500:1
1000:1
When an active probe is inserted, the following menus are added in addition to the
Probe Type and Ratio menus.
1. Front-End
Press this softkey to select the probe front end that matches the probe actually
connected. You can select single-end or differential.
2. Probe-Cal
The oscilloscope provides the calibration function for active probes. This function
uses the fast edge signal output from the [Trig Out/Calibration] connector on
the rear panel of the instrument as the probe calibration signal. Before
performing probe calibration, press UTIL AuxOutput and select "Fast".
The following part introduces how to calibrate the probe via the oscilloscope
(take CH1 as an example) using PCK100 active differential probe calibration kit
by taking RP7150 as an example. For the detailed calibration procedures, please
refer to
RP7000 Series Active Probe User’s Guide
.
1) Connect RP7150 to CH1 input terminal of the oscilloscope correctly.
2) Press CH1 Probe Probe-Cal to open the probe calibration menu. At

Table of Contents

Other manuals for Rigol DS6104

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS6104 and is the answer not in the manual?

Rigol DS6104 Specifications

General IconGeneral
BrandRigol
ModelDS6104
CategoryTest Equipment
LanguageEnglish

Related product manuals