3 To Set the Horizontal System RIGOL
DS6000 User’s Guide 3-3
The waveform before enlargement:
The waveform in the area that has not been covered by the subtransparent blue in
the upper part of the screen is the waveform before enlargement. You can turn
HORIZONTAL
POSITION to move the area left and right or turn
HORIZONTAL
SCALE to enlarge or reduce this area.
The waveform after enlargement:
The waveform in the lower part of the screen is the horizontally expanded waveform.
Note: The delayed sweep time base has increased its resolution relative to the main
time base (as shown in the figure above). The delayed sweep time base should be
lower than or equal to the main time base.
Tip
When the channel currently selected is CH1 to CH4, you can turn on or off the
delayed sweep not only through MENU in the horizontal control area
(HORIZONTAL) on the front panel but also by directly pressing HORIZONTAL
SCALE
(the delayed sweep shortcut key) to switch the on/off status of the delayed
sweep.