CAN serial bus trigger settings panel
NOTE. Requires option SRAUTO.
Field or control Description
Source Select the CAN bus on which you want to trigger.
Trigger On Select the type of information on which to trigger.
Frame Type Sets the frame type on which to trigger.
Available when Trigger On = Type of Frame.
Direction Sets the transfer direction on which to trigger (Read, Write, Either).
Available when Trigger On = Identifier.
Identifier Format Sets the identifier for standard (11-bit) or extended (29-bit for CAN 2.0B) length.
Available when Trigger On = Identifier or ID & Data.
Identifier Sets the identifier pattern on which to trigger. The number of bits shown depends on the
Identifier Format setting.
Tap the Binary or Hex field and use the A and B knobs to select and change the values. Or
double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Identifier or ID & Data.
Data Bytes Sets the number of data bytes on which to trigger (one to eight bytes). Use the A knob to
change the value.
Available when Trigger On = Data or ID & Data.
Data Sets the data pattern on which to trigger. The number of bits shown depends on the Data Bytes
setting.
Tap the Binary or Hex field and use the A and B knobs to select and change the values. Or
double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Data or ID & Data.
Data Offset Sets the byte offset value.
Available when Trigger On = Data or ID & Data.
Trigger When Sets the trigger when condition.
Available when Trigger On = Data or ID & Data.
Bit Rate Switch (BRS) Bit Sets the BRS bit state on which to trigger.
Available when Source is a CAN FD bus and Trigger On = FD Bits.
Error Status Indicator (ESI)
Bit
Sets the ESI bit state on which to trigger.
Available when Source is a CAN FD bus and Trigger On = FD Bits.
Menus and dialog boxes
348 MSO54, MSO56, MSO58, MSO58LP, MSO64 Help

Table of Contents

Other manuals for Tektronix MSO54B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54B and is the answer not in the manual?

Tektronix MSO54B Specifications

General IconGeneral
BrandTektronix
ModelMSO54B
CategoryTest Equipment
LanguageEnglish

Related product manuals