FlexRay serial bus trigger settings panel
NOTE. Requires option SRAUTO.
Field or control Description
Source Select the FlexRay bus on which to trigger.
Trigger On Select the type of information on which to trigger.
Indicator Bits Select the indicator bits type on which to trigger.
Tap the Binary or Hex field and use the A and B knobs to select and change the values. Or
double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Indicator Bits or Header Fields.
Identifier Sets the frame identifier pattern on which to trigger.
Tap the Binary, Hex, or Decimal field and use the A and B knobs to select and change the
values. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Frame ID, Header Fields, or Identifier & Data.
Cycle Count Sets the cycle count pattern on which to trigger. Use in conjunction with the Trigger When field
to specify the exact trigger condition.
Tap the Binary, Hex, or Decimal field and use the A and B knobs to select and change the
values. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Cycle Count or Header Fields.
Payload Length Tap the Binary, Hex, or Decimal field and use the A and B knobs to select and change the
values. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Header Fields.
Header CRC Tap the Binary or Hex field and use the A and B knobs to select and change the values. Or
double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Header Fields.
Data Bytes Sets the number of data bytes on which to trigger (one to sixteen bytes). Use the A knob to
change the value. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Data or Identifier & Data.
Data Sets the data pattern on which to trigger. The number of bits shown depends on the Data Bytes
setting. Use in conjunction with the Trigger When field to specify the exact trigger condition.
Tap the Binary or Hex field and use the A and B knobs to select and change the values. Or
double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Data or Identifier & Data.
Byte Offset Sets the byte offset (Don't Care or the number of bytes). Tap the input box and use the A knob
to change the value.
Available when Trigger On = Data or Identifier & Data.
Menus and dialog boxes
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help 353