Settings panel (Runt Trigger configuration menu) fields and controls.
Field or control Description
Source Lists the source channel or waveform to use to trigger or search. Types that require multiple
inputs will replace this control with a different source definition control.
Trigger When
â–
Occurs: A runt signal event occurs.
â–
< Limit: A runt signal event occurs that has a pulse width less than the specified time limit.
â–
> Limit: A runt signal event occurs that has a pulse width greater than the specified time
limit.
â–
= Limit: A runt signal event occurs that has a pulse width that is equal to the specified time
limit (±5%).
â–
≠Limit: A runt signal event occurs that has a pulse width that does not equal (is greater
than or less than) the specified time limit (±5%).
Polarity Sets the polarity of the pulse to detect (positive pulse only, negative pulse only, or a positive or
negative pulse).
Time Limit Sets the time period condition to be met.
Upper Threshold Sets the upper threshold amplitude level through which the signal must pass to be considered a
valid transition.
Lower Threshold Sets the lower threshold amplitude level through which the signal must pass to be considered a
valid transition.
Logic Qualification Set to On to enable logic qualification to further refine the trigger condition by setting the
required logic conditions on the source signals to generate a trigger event.
Define Inputs Opens the Logic Qualification - Define Inputs configuration menu. Use this menu to set the
logic state, threshold levels, and logic operation of the input signals. See Logic Qualification -
Define Inputs configuration menu on page 400.
Available when Logic Qualification = On.
Menus and dialog boxes
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help 389