IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
Page 8-22 Section 8-8 - Rear Panel
13.) For assembly follow the reverse order.
Table 8-10 Functional Tests
Section 8-8 Rear Panel
8-8-1 Disassembly/Assembly of Rear Panel
Follow steps below and refer 8-14
1.) Remove Top Cover. Refer 8-2. For FRU Details refer
2.) Remove Noise filter by removing screws (c,d) then disconnect the power cables from the noise
filter.
3.) Remove all five screws indicated in the fig below from the rear panel.
4.) Remove the connectors from the Davinci board, FEB and CRT PCB.
Figure 8-13 Disassembly/Assembly of Power Supply/Distributor PCB
Service Manual
Section Functional Test/ Diagnostic Test Debrief Script
4-2-1 Power ON/ Boot up
Service manual direction 5305880-
100, Section 8-7-1.Equipment
passed all required tests and is
ready for use.
4-2-1 Power OFF/ Shut down
4-3 Functional Checks
4-5 Software configuration check
7-2 Diagnostics tests