1-5-5 General Notice
1.) Designation of Peripheral Equipment Connectable to this Product. The peripheral which
conforms to EN60601-1-2 (IEC601-1-2), can be connected up to the LOGIQ™ 100 PRO
without compromising its EMC performance. Failure to comply with the above instruction may
result in poor EMC performance of the product.
2.) Notice against User Modification: Do not modify this product. Unilateral user modification
may cause degradation in EMC performance. Modification of the product includes:
a.) Changes in cables (length, material, wiring etc.)
b.) Changes in system installation/layout
c.) Changes in system configuration/components
d.) Changes in means of fixing system/parts (cover open/close, cover screwing)
3.) Operate the system with all covers closed. If you open any cover for some reason, be sure to
shut it before starting/ resuming operation. Operating the system with any cover open may
affect EMC performance.
1-5-6 Countermeasures against EMC related Issues
Generally it is very difficult to handle with issues related to EMC. It is time consuming and
expensive. General counter measures for Electromagnetic interference with other equipment
1.) Electromagnetic interference may be reduced by positioning other equipment far away from
the system.
2.) Electromagnetic interference may be reduced by changing the relative location (installation
angle) of the system and other equipment.
3.) Electromagnetic interference may be reduced by changing wiring locations of power/signal
cables of other equip ment.
4.) Electromagnetic influence may be reduced by altering the path of power supply for other
equipment.
1-5-7 Notice on Service
Ensure all screws are tight after servicing. Loose screws may cause degradation in EMC
performance.
CAUTION Do not use the following devices near the LOGIQ™ 100 PRO system. Devices which
intrinsically transmit radio waves such as cellular phone, radio transceiver, mobile radio
transmitter radio-controlled toy, etc. Use of these devices could cause the LOGIQ™ 100 PRO
system to perform outside the published specifications. Keep power to these devices turned
OFF when near the system. Medical staff in charge of the LOGIQ™ 100 PRO system is
required to instruct technicians, patients and other people who may be around the system
to fully comply with the above regulation
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 37 of 197

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 100 PRO and is the answer not in the manual?

GE LOGIQ 100 PRO Specifications

General IconGeneral
BrandGE
ModelLOGIQ 100 PRO
CategoryMedical Equipment
LanguageEnglish

Related product manuals