GE HEALTHCARE
D
IRECTION 5305880-100, REVISION 3 LOGIQ™ 100 PRO SERVICE MANUAL
Page 10-18 Section 10-6 - Electrical Safety Tests
10-6-5-4 Data Sheet for Chassis Source Leakage Current
The test passes when all readings measure less than the value shown in Table 10-13. Record all
data on the PM Inspection Certificate.
Unit
Power
Tester
Polarity
Switch
Tester Neutral
or Ground
Switch
Test 1 Probe
Connector
Screw
Test 2
Rear
Panel
Optional
Test 4
Optional
Test 5
Enter Name of tested peripheral here:
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED
Table 10-17 Typical Data Sheet for Chassis Source Leakage Current
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 189 of 197