GE MEDICAL SYSTEMS
D
IRECTION 2300000, REVISION 1 LOGIQ™ 5 SERVICE MANUAL
Chapter 5 Components and Functions 5 - 27
5-8-3 DC Power (Low Voltage)
DC Power are supplied by LV unit in AC Power assy. DC Power of LV unit are power source of
all assys in nest and nest fan, etc
JPC assy in AC power assy is DC power source also. DC power of JPC is for on/off control circuit
of VIC assy in BEP assy
DC Output Capacity:
• +3.3V, 5.6A
•+5V, 7A
• +6V, 8.7A
• +6V, 8.7A
• -6V, 3.4A
• +15V, 8A
• -15V, 1A