GE MEDICAL SYSTEMS
DIRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
4-4 Section 4-3 - General Procedure
4-3-2-2 Switch off the System
1.) Switch OFF the Circuit Breaker at the back of the scanner.
2.) Disconnect the Mains Power Cable is necessary. For example: Servicing or relocating the scanner.
4-3-3 System Features
4-3-3-1 Control Panel
1.) Touch Panel and Controls
2.) Additional Touch Panel and Controls
3.) Touch Panel Brightness Control
4.) Video
5.) Audio On/Off and Volume
6.) TGC
7.) Reverse
8.) Patient Button
9.) Alphanumeric Keyboard
NOTICE
Disconnecting the Mains Power Cable before switching OFF the Circuit Breaker will activate the
uninterruptible power system (UPS) in the Back End Processor, forcing an ordered shutdown of the
system.
Figure 4-4 Control Panel Tour
PDI
PW
L
R
B Pause
Auto
Depth
M
Zoom
Ellipse
Body Pattern
Measure
M/D Cursor
Scan Area
Set
Clear
Comment
Patient
2
1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17