GE MEDICAL SYSTEMS
DIRECTION 2300000, REVISION 1 LOGIQ™ 5 SERVICE MANUAL
6 - 4 Section 6-3 - Power Supply Adjustment
6-3-2 Adjustments Procedures
1.) Using the following VRs, DC output can be adjusted.
Table 6-20 DC Output Specification for LV Unit
VR
Specifications
VR1
6V
VR2
3.3V
VR3
6V
VR4
-6V
VR5
-15 V
VR6
5V
VR7
15V
Figure 6-3 VRs for Adjusting DC Output
VR1
VR2
VR3
VR4
VR5
VR6
VR7