7-4-1 Check and Record the P4 Key Function (cont’d)
5.) In the Destinations section, record the service that is displayed. The destinations list displays
the following information:
* Name: user defined during service configuration
* Type: the type of service
* Server: the device for which the service was configured
* Dir: direction: output, input, or both (I+O)
6.) In the Image generated section, record the parameters related to the service.
7-4-2 Setting the P4 Key to Screen Capture
If the P4 Key is not set to screen capture:
1.) While on the Connectivity screen, with the Buttons tab displayed, go to the drop down selection
menu in the Destinations section.
2.) From the drop down menu select CopyToWflow_01>Image to Buffer>MyComputer>Out.
3.) Ensure that the Image generated section for capture Area is set to Whole Screen, secondary
Capture and No Image Compression.
4.) The P4 Key should now be set up for whole screen capture, sending the screens to the image
buffer (clipboard).