General procedures
LOGIQ e – Basic Service Manual 4-11
5461614-100 English Rev. 6
Power shut down
When you switch off the unit, the system performs an automatic
shutdown sequence.
1. Press the Power On/Off switch at the front of the system
once.
2. The System-Exit window is displayed.
Figure 4-5. System Exit menu
3. Using the Trackball or Select key, select Shutdown.
4. The shutdown process takes 15 seconds and the power off
sequence is complete when the power status LED is turned
off.
5. Disconnect the probes. Clean or disinfect all probes as
necessary. Store them in their shipping cases to avoid
damage.
6. Close LCD cover.