EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ e

GE LOGIQ e Basic Service Manual

GE LOGIQ e
427 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #301 background imageLoading...
Page #301 background image
System maintenance
LOGIQ e – Basic Service Manual 10-17
5461614-100 English Rev. 6
System maintenance
Contents in This Section
• ‘Preliminary checks’ on page 10-17
• ‘Functional checks’ on page 10-19
• ‘System checks’ on page 10-19
• ‘Peripheral/option checks’ on page 10-20
• ‘Mains cable inspection’ on page 10-20
• ‘Cleaning’ on page 10-21
• ‘Physical inspection’ on page 10-21
• ‘Optional Diagnostic Checks’ on page 10-22
• ‘Probe maintenance’ on page 10-23
• ‘Basic probe care’ on page 10-23
• ‘Basic probe cleaning’ on page 10-24
• ‘Battery Performance Maintenance’ on page 10-25
Preliminary checks
The preliminary checks take about 15 minutes to perform. Refer
to the Ultrasound system user documentation whenever
necessary.
Table 10-7: System preliminary checks
Step Item Description
1. Ask and Listen Ask the customer if they have any problems or questions about the
equipment.
2. Paperwork Fill in the top of Ultrasound Inspection Certificate (see Figure 10-10 on
page 10-40). Record all probes and Ultrasound system options.
3. Power up • Turn the Ultrasound system power on and verify that all fans and
peripherals turn on.
• Watch the displays during power up to verify that no warning or error
messages are displayed.
• Where applicable, confirm that the battery is charged. If no AC Input
present, use the internal battery.
4. Probes Verify that the Ultrasound system properly recognizes all probes.

Table of Contents

Other manuals for GE LOGIQ e

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ e and is the answer not in the manual?

GE LOGIQ e Specifications

General IconGeneral
BrandGE
ModelLOGIQ e
CategoryMedical Equipment
LanguageEnglish

Related product manuals