Care & Maintenance
LOGIQ e – Basic Service Manual 12-29
5461614-100 English Rev. 6
Cleaning Dusk Screen
The dusk screen requires weekly cleaning in order to retain its
performance. Power off the Power Box when the dusk screen is
removed for cleaning.
1. Pull out the Dusk Screen from the power box.
Figure 12-2. Pull out the Dusk Screen
2. Dust the Dusk Screen with a vacuum cleaner and/or wash it
with a mild soapy solution. If washed, rinse and dry the Dusk
Screen before re-installation.
Figure 12-3. Clean the Dusk Screen
3. Re-install the Dusk Screen on the power box.
Figure 12-4. Re-install the Dusk Screen