EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ e

GE LOGIQ e Basic Service Manual

GE LOGIQ e
427 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #150 background imageLoading...
Page #150 background image
General Procedures and Functional Checks
4-16 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Lockout/Tagout (LOTO) requirements
Follow OSHA Lockout/Tagout requirements (USA) or local
Lockout/Tagout requirements by ensuring you are in total
control of the AC power plug at all times during the service
process.
To apply Lockout/Tagout:
1. Plan and prepare for shutdown.
2. Shutdown the equipment.
3. Isolate the equipment.
4. Apply Lockout/Tagout Devices.
5. Remove battery.
6. Control all stored and residual energy.
7. Verify isolation.
All potentially hazardous stored or residual energy is
relieved.
WARNING
Energy Control and Power Lockout for LOGIQ e
When servicing parts of the system where there is exposure to
voltage greater than 30 volts:
1. Turn off the scanner.
2. Unplug the system.
3. Maintain control of the system power plug.
4. Wait for at least 20 seconds for capacitors to discharge as
there are no test points to verify isolation. The amber light
on the operator panel On/Off button will turn off.
5. Remove the system battery.
All potentially hazardous stored or residual energy is relieved.

Table of Contents

Other manuals for GE LOGIQ e

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ e and is the answer not in the manual?

GE LOGIQ e Specifications

General IconGeneral
BrandGE
ModelLOGIQ e
CategoryMedical Equipment
LanguageEnglish

Related product manuals