Docking Cart Setup
11-40 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Cart Using
Contents in This Section
• ‘Introduction’ on page 11-40
• ‘Height Adjustment’ on page 11-40
• ‘Locking the Wheels’ on page 11-41
• ‘Mounting the System to Cart’ on page 11-41
• ‘Release the System from Docking Cart’ on page 11-42
• ‘Switch the Three Probes’ on page 11-43
• ‘System Operation’ on page 11-43
Introduction
Docking Cart supports the ultrasound system as below:
• LOGIQ e (Software version R8.x.x)
The docking cart can supply power for the system and
peripherals which are mounted to the docking cart.
Height Adjustment
To adjust the height of the mounting-platform, raise the Release
lever and pull the platform up or push it down with both hands,
the traveling distance is 140mm.
When the lever is released the platform remains at the adjusted
height.