Replacement Procedures
8-16 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Loading the System Software
NOTE: Before loading the system software, please ensure that the
power can be continuously supplied and there is no risk of
power cut off during loading procedure.
There are two methods to load the system software:
Load the system software with USB memory stick.
Loading the System Software with USB memory stick
NOTE: Before starting this procedure, remove all probes and
peripherals and remove them from the Docking Cart.
NOTE: While it is believed to be unnecessary, it would not hurt to
disconnect the system from the network and remove all
transducer.
NOTE: Please ensure AC adapter is connected during system upgrade!
1. Insert the USB memory stick labeled “System & Application
Software” to the system.
2. Properly turn off the scanner by momentarily pressing the
Power On/Off Switch. In System-Exit window, select
Shutdown to shutdown the system.
NOTE: If the system will not shutdown normally, hold down the
Power On/Off Switch until the light turns off.
3. Power on the system. The system will detect the USB
memory stick automatically.
WARNING
While the software install procedure is designed to
preserved data, you should save any patient data,
images, system setups and customer presets to CD,
DVD, USB Flash Drive, or USB Hard Disk before doing a
software upgrade.

Table of Contents

Other manuals for GE LOGIQ e

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ e and is the answer not in the manual?

GE LOGIQ e Specifications

General IconGeneral
BrandGE
ModelLOGIQ e
CategoryMedical Equipment
LanguageEnglish

Related product manuals