USB Quick Save
LOGIQ e – Basic Service Manual 7-7
5461614-100 English Rev. 6
USB Quick Save
Overview
There may be times when the customer or field engineer will
want to directly save images into USB memory. This is
accomplished by saving individual Cine clips (moving images-
avi format) or still images (jpg format) directly to a USB memory
disk by pressing a Print Key.
The P3 key is the factory default print key to accomplish the
USB Quick Save. However, the default is for the Image Area
only or the customer may have customized the default Key
function.
Check and Record the P3 Key Function
Check the function of the default Key in the event that the
customer may have made some customized settings.
1. Click Utility on the keyboard.
2. Select Connectivity from the Utilities Menu.
3. Select the Button tab on the Connectivity screen.
4. In the Physical Print Buttons section, select Print3 key.
The Connectivity/Buttons Screen will be displayed.
Figure 7-3. Buttons Set Up Screen