Loading the software
LOGIQ e – Basic Service Manual 8-13
5461614-100 English Rev. 6
Loading the software
Contents in This Section
• ‘Purpose of this section’ on page 8-13
• ‘Customer provided prerequisite’ on page 8-13
• ‘Data Management - moving all images’ on page 8-14
• ‘Backing up the Patient Archive and System Configurations’
on page 8-14
• ‘Recording important settings and parameters’ on page 8-15
• ‘Loading the System Software’ on page 8-16
• ‘Loading the System Software with USB memory stick’ on
page 8-16
• ‘Software Version check out’ on page 8-21
• ‘Reload the Correct Preset Region’ on page 8-22
• ‘Option Strings Check’ on page 8-23
• ‘Probe Recognition Check’ on page 8-24
• ‘Peripheral Device Check’ on page 8-24
• ‘Reinstall DICOM Devices’ on page 8-25
Purpose of this section
This section describes how to reinstall software on LOGIQ e.
Customer provided prerequisite
• Formatted and labelled media for Images storage.
• Formatted and labelled media for Patient Archive and
Presets (User Defined Settings).
• Password for the user ADM.