4.18 Patient Pendant
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 56 CareAssist® and CareAssist® ES Bed Service Manual (MAN330 REV 4)
4.18 Patient Pendant
Tools required: Small blade screwdriver
Removal
1. Set the brakes.
2. Raise the sleep surface to the highest position.
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Disconnect the pendant cable from pendant mount on the bed.
5. Remove the pendant from the bed.
Replacement
1. Make sure the new pendant has the same functions as the pendant being
replaced.
2. Perform the removal procedure in reverse order.
3. To make sure the CareAssist® Bed operates correctly, do the “Function
Checks” on page 2-5.