2.10 CPR Malfunction
Chapter 2: Troubleshooting Procedures
CareAssist® and CareAssist® ES Bed Service Manual (MAN330 REV 4) Page 2 - 27
2
2.10 CPR Malfunction
Initial conditions: the hilow function works (see “Function Checks” on page
2-5).
1. Check the mechanics of the CPR system:
Remove the seat, knee and foot sleep deck sections. Make sure the
mounting hardware is not damaged or missing from the release mechanism
of the head section motor.
All hardware is secure and not damaged.
Yes No
↓→Replace the missing or damaged parts, and then go to “Function
Checks” on page 2-5.
2. Check that the release cable is not broken or damaged.
The cable is intact.
Yes No
↓→Go to step 4.
3. Check that the adjustment of the release cable is correct (refer to procedure
4.6 on page 4-20).
The adjustment is correct.
Yes No
↓→Adjust the release cable (refer to procedure 4.6 on page 4-20).
4. Replace the head section motor (refer to procedure 4.5 on page 4-15) and
release cable subassembly (refer to procedure 4.6 on page 4-20).
This solves the problem.
Yes No
↓→Contact Hill-Rom technical support.
5. Go to “Final Actions” on page 2-12.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom CareAssist P1170 and is the answer not in the manual?

Hill-Rom CareAssist P1170 Specifications

General IconGeneral
BrandHill-Rom
ModelCareAssist P1170
CategoryMedical Equipment
LanguageEnglish

Related product manuals