2.2 Lockout Malfunction
Chapter 2: Troubleshooting Procedures
Page 2 - 16 CareAssist® and CareAssist® ES Bed Service Manual (MAN330 REV 4)
2.2 Lockout Malfunction
NOTE:
The symptom of this fault is after pressing the lockout button, none of the
indicators flash.
NOTE:
The Bed Not Down indicator does not flash in the steps that include the
asterisk symbol (*).
1. At least one of the other functions operate correctly.
Yes No
↓→Go to RAP 2.1.
2. Verify the correct connection in the interface box:
a. Swap siderail cable connections inside the interface box (A model
beds) or inside the power supply (B model and newer beds).
b. Press the Lockout control.
All the indicators flash (*).
Yes No
↓→Go to step 4.
3. Replace the siderail interface P.C. board (refer to procedure 4.15) (A model
beds) or (refer to procedure 4.31) (B model and newer beds) on the affected
siderail.
4. Do the following to make sure the caregiver siderail controls operate
properly:
a. Disconnect the control unit.
b. Connect a known good part.
c. Press the function on the caregiver control panel.
All the LEDs flash (*).
Yes No
↓→Replace the siderail control panel (refer to procedure 4.17 on
page 4-53) with the proper configuration. If this solves the
problem perform the “Final Actions” on page 2-12, otherwise
contact Hill-Rom Technical Support.
5. Go to “Final Actions” on page 2-12.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom CareAssist P1170 and is the answer not in the manual?

Hill-Rom CareAssist P1170 Specifications

General IconGeneral
BrandHill-Rom
ModelCareAssist P1170
CategoryMedical Equipment
LanguageEnglish

Related product manuals