4.19 Patient Pendant Mount
Chapter 4: Removal, Replacement, and Adjustment Procedures
CareAssist® and CareAssist® ES Bed Service Manual (MAN330 REV 4) Page 4 - 57
4
4.19 Patient Pendant Mount
Tools required: Wire cutters
This procedure is for moving the pendant mount from one side of the bed to the
other side of the bed.
Removal
1. Set the brakes.
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
2. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
3. Remove the pendant (refer to procedure 4.18).
4. Cut and remove the cable ties securing the pendant mount to the bed.
Replacement
1. Move the pendant mount to the opposite side of the bed.
2. Using cable ties, secure the pendant mount to the bed.
3. To make sure the CareAssist® Bed operates correctly, do the “Function
Checks” on page 2-5.