2.8 Trendelenburg/Reverse Trendelenburg Malfunction
Chapter 2: Troubleshooting Procedures
CareAssist® and CareAssist® ES Bed Service Manual (MAN330 REV 4) Page 2 - 23
2
2.8 Trendelenburg/Reverse Trendelenburg Malfunction
Initial conditions: the hilow function works (see “Function Checks” on page
2-5).
1. At least one of the other functions is working.
Yes No
↓→Go to RAP 2.1.
2. Do the following to make sure the caregiver controls operate correctly.
a. Disconnect the control unit.
b. Connect a known good control unit.
The trendelenburg/reverse trendelenburg function works.
Yes No
↓→Replace the power supply (A model beds) (refer to procedure
4.3) or the motor control P.C. board (B model and newer beds)
(refer to procedure 4.27). If this solves the problem perform the
“Final Actions” on page 2-12, otherwise contact Hill-Rom
Technical Support.
3. Replace the affected control unit.
This solves the problem.
Yes No
↓→Contact Hill-Rom Technical Support.
4. Go to “Final Actions” on page 2-12.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom CareAssist P1170 and is the answer not in the manual?

Hill-Rom CareAssist P1170 Specifications

General IconGeneral
BrandHill-Rom
ModelCareAssist P1170
CategoryMedical Equipment
LanguageEnglish

Related product manuals