Electromagnetic Emissions and Immunity Guidance
The CareAssist® and CareAssist® ES Bed are intended for use in the
electromagnetic environment specified in the tables below. The customer or
the user of the bed should make sure that it is used in such an environment
Table 1-9. Electromagnetic Emissions Guidance
Table 1-10. Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The CareAssist® and CareAssist® ES Bed model P1170 is intended for use in the electromagnetic environment specified
below. The customer or the user of the model P1170 should make sure it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment—Guidance
RF emissions
CISPR 11
Group 1 The model P1170 uses RF energy only for its internal
functions. Therefore, its RF emissions are low and are
not likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class A The model P1170 is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Not Applicable
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Not Applicable
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The CareAssist® and CareAssist® ES Bed model P1170 is intended for use in the electromagnetic environment specified
below. The customer or the user of the model P1170 should make sure it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment—Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Radiated RF
IEC 61000-4-3
3 Vrms
80 MHz to 2.5 GHz
3 Vrms
80 MHz to 2.5 GHz
Portable and mobile RF communica-
tions equipment should not be used
at close distances to the P1170 bed.
(See Note 2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom CareAssist P1170 and is the answer not in the manual?

Hill-Rom CareAssist P1170 Specifications

General IconGeneral
BrandHill-Rom
ModelCareAssist P1170
CategoryMedical Equipment
LanguageEnglish

Related product manuals