SCPI Command Reference 17-37
Composite testing
PASS:SOURce2 <NRf> | NDN
:CALCulate2:CLIMits:PASS:SOURce2 <NRf> | <NDN> Specify composite pass pattern
Parameters <NRf> = 0 to 7 (3-bit) Decimal value
0 to 15 (4-bit) Decimal value
<NDN> = 0 to #b111 (3-bit) Binary value
0 to #b1111 (4-bit) Binary value
0 to #q7 (3-bit) Octal value
0 to #q17 (4-bit) Octal value
0 to #h7 (3-bit) Hexadecimal value
0 to #hF (4-bit) Hexadecimal value
Query :SOURce2? Query programmed source value
Description This command is used to dene the 3-bit or 4-bit output pattern for the Dig-
ital I/O Port when there are no failures. Note that the output value can be
specied in binary, octal, decimal, or hexadecimal format. Use the table pro-
vided in :SOURce2<NRf> | <NDN>, Description to determine the parame-
ter value for the desired decimal digital output pattern.
The SourceMeter can be congured to place the dened pass bit pattern
on the digital output immediately when the pass condition occurs, or it
can wait until all testing on a device package is completed (operation leaves
trigger layer). See :BCONtrol for details.
For the sorting mode, this command denes the 3-bit or 4-bit output pass
pattern for the Limit 1 test (compliance) when Limits 2, 3, and 5-12 are
disabled.
NOTE 16-bit I/O is available with the 2499-DIGIO option. The maximum 16-bit output
value is 65535.
FAIL:SOURce2 <NRf> | <NDN>
:CALCulate2:CLIMits:FAIL:SOURce2 <NRf> | <NDN> Specify fail pattern
Parameters <NRf> = 0 to 7 (3-bit) Decimal value
0 to 15 (4-bit) Decimal value
<NDN> = 0 to #b111 (3-bit) Binary value
0 to #b1111 (4-bit) Binary value
0 to #q7 (3-bit) Octal value
0 to #q17 (4-bit) Octal value
0 to #h7 (3-bit) Hexadecimal value
0 to #hF (4-bit) Hexadecimal value
Query :SOURce2? Query programmed source value
Description For the sorting mode, this command is used to dene the 3-bit or 4-bit
output pattern for the Digital I/O Port when there are failures. Note that the
output value can be specied using binary, octal, decimal, or hexadecimal
format. Use the table provided in :SOURce2<NRf> | <NDN>, Description
to determine the decimal parameter value for the desired digital output
pattern.

Table of Contents

Other manuals for Keithley 6430

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
TypeSub-Femtoamp Remote SourceMeter
Compliance Voltage200 V
Maximum Output Voltage200 V
Maximum Output Current100 mA
Compliance LimitProgrammable
InterfaceGPIB, RS-232
Input Impedance>10^14 Ω

Summary

Safety Precautions

3 Basic Source-Measure Operation

Basic Source-Measure Procedure

Details setup for source-measure operations: function, values, limits, and output control.

5 Source-Measure Concepts

Compliance limit

Discusses compliance limits, values, and determination.

Operating boundaries

Covers voltage/current boundaries for source, sink, I-source, V-source, and source-measure modes.

6 Range, Digits, Speed, and Filters

Range and digits

Discusses available ranges, limits, manual/autoranging, and display resolution.

Filters

Provides information on 3-stage filtering to reduce reading noise.

7 Relative and Math

Math Operations

Details math (FCTN) operations: power, offset ohms, varistor, etc.

9 Sweep Operation

Configuring and Running a Sweep

Discusses setup, configuration, delay setting, and performing sweeps.

10 Triggering

Trigger Model

Discusses trigger model, layers, event detection, delay, and device action.

Configuring Triggering

Details how to configure various triggering aspects.

Remote Triggering

Details remote trigger model, commands, and basic triggering example.

11 Limit Testing

Types of Limits

Discusses three limit types: compliance, coarse, fine; summarizes grading and sorting modes.

Configuring and Performing Limit Tests

Describes configuration for limit tests and typical test procedure.

12 Digital I/O Port, Interlock, and Output Configuration

Safety Interlock

Describes using Digital I/O Port as a safety interlock.

14 Status Structure

Status Byte and Service Request (SRQ)

Explains programming Status Byte for SRQs and using serial poll to detect SRQs.

15 Common Commands

Command Summary

Lists IEEE-488.2 common commands used by the SourceMeter.

Command Reference

Provides detailed reference for common commands, excluding status structure.

16 SCPI Signal-Oriented Measurement Commands

Configuring Measurement Function

Provides detailed info on commands to configure measurement function.

Acquiring Readings

Describes commands to acquire post-processed readings, trigger, acquire, and single measurement.

17 SCPI Command Reference

18 Performance Verification

Performing the Verification Test Procedures

Details restoring factory defaults, setting ranges, and output values.

Mainframe Verification

Covers procedures to verify mainframe accuracy alone.

19 Calibration

Mainframe Calibration

Includes procedures to calibrate the mainframe without the Remote PreAmp.

20 Routine Maintenance

Line Fuse Replacement

Covers procedure and part numbers for replacing the line fuse.

F Measurement Considerations

Floating measurement safety concerns

Discusses safety concerns for measurements up to 42V above chassis ground.

Leakage currents and guarding

Explains reducing leakage currents using insulators, humidity control, and guarding.