9-22 Sweep Operation
Source memory sweep programming example
As an example of source memory sweep operation, assume a three-point sweep with the fol-
lowing operating modes:
Source Memory Location #1: source voltage, measure current, 10V source value
Source Memory Location #2: source current, measure voltage, 100mA source value
Source Memory Location #3: source current, measure current, 100mA source value
Table 9-8 summarizes the basic remote command sequence for performing the basic source
memory sweep described above.
Table 9-8
Source memory sweep programming example
Command Description
*RST
:SENS:FUNC:CONC OFF
:SOUR:FUNC MEM
:SOUR:MEM:POIN 3
:SOUR:MEM:STAR 1
:SOUR:FUNC VOLT
:SENS:FUNC ‘CURR:DC’
:SOUR:VOLT 10
:SOUR:MEM:SAVE 1
:SOUR:FUNC CURR
:SENS:FUNC ‘VOLT:DC’
:SOUR:CURR 100E-3
:SOUR:MEM:SAVE 2
:SENS:FUNC ‘CURR:DC’
:SOUR:MEM:SAVE 3
:TRIG:COUN 3
:OUTP ON
:READ?
Restore GPIB default conditions.
Turn off concurrent functions.
Source memory sweep mode.
Number memory points = 3.
Start at memory location 1.
Volts source function.
Current sense function.
10V source voltage.
Save in source memory location 1.
Current source function.
Volts sense function.
100mA source current.
Save in source memory location 2.
Current sense function.
Save in source memory location 3.
Trigger count = # sweep points.
Turn on source output.
Trigger sweep, request data.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals