A-12 Specifications
Cases V and VI: Measure one function, Output Auto-Off disabled, and no
source setting changes.
Auto-Zero: Enabled
Trigger Latency: 500µsec max
Auto Zero: Disabled
Trigger Latency: 100µsec max
The source turns on as soon as the output is turned on and remains on until the source is
turned off. As shown in the two timing diagrams, the static source remains on for every mea-
surement cycle. The Source-Delay portion of the SDM cycle is omitted. With Trigger Delay set
to zero, Trigger Latency is the time from when the trigger event occurs to when the Source-
Meter begins an A/D conversion.
Trigger
Latency
Trigger
Delay
Trigger
Event
Source On
A/D
Conversion
(signal phase)
A/D
Conversion
(ref phase)
A/D
Conversion
(ref zero phase)
F
i
g
u
r
e A-5
Case V timing
diagram
Trigger
Latency
Trigger
Event
Source On
Trigger
Delay
A/D
Conversion
(signal phase)
F
i
g
u
r
e A-6
Case VI timing
diagram