EasyManuals Logo
Home>Keithley>Measuring Instruments>6430

Keithley 6430 User Manual

Keithley 6430
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #472 background imageLoading...
Page #472 background image
19-22 Calibration
Viewing calibration dates and calibration count
When calibration is locked, only the UNLOCK and VIEW-DATES selections will be acces-
sible in the mainframe calibration menu. To view calibration dates and calibration count:
1. From normal display, press MENU, select CAL, and then press ENTER. The unit will
display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
2. Select VIEW-DATES, and then press ENTER. The Model 6430 will display the next
and last calibration dates and the calibration count as in the following example:
NEXT CAL: 12/15/1999
Last calibration: 12/15/1999 Count: 0001

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals