18-30 Performance Verification
4. Verify ohms measurement accuracy for each of the resistance values listed in
Table 18-15. For each measurement:
Connect the BNC shorting cap to select the appropriate resistance value.
Select the appropriate ohms measurement range with the RANGE keys.
Verify that the Model 6430 resistance reading is within the limits given in the
table.
2T and 20T range accuracy
1. With the power off, connect the characterized 1T resistor directly to the Remote
PreAmp IN/OUT HIGH jack as shown in Figure 18-10. (Use a male triax-to-triax
adapter.)
Table 18-15
Remote PreAmp 2G-200G range measurement accuracy limits
Model 6430 range Standard resistance
1
Model 6430 resistance reading limits
2
(1 year, 18°-28°C)
2G
20G
200G
1G
10G
100G
0.99875 to 1.00125G
9.9865 to 10.0135G
99.735 to 100.265G
1
Nominal values shown. Use characterized values.
2
Calculated using Model 6430 normal accuracy specications and Model 5156 nominal values. Limits include
Model 5156 characterization accuracy.
Model 6430
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V, ,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
Remote PreAmp
Mainframe/PreAmp
Cable
Connect Resistor to
IN/OUT HIGH Jack
using Triax Adapter
Test Resistor
(See Figure 18-1)
250
PEAK
40V
PEAK
GUARD
IN/OUT
LOW
CAT I
MADE IN
U.S.A.
IN/OUT
HIGH
SENSE
MAINFRAME
250
PEAK
42V
PEAK
40V
PEAK
250
PEAK
WARNING: NO INTERNAL OPERATOR SERVICEABLE
PARTS SERVICE BY QUALIFIED
PERSONNEL ONLY
!
Figure 18-10
Connections for Remote
PreAmp 2T and 20T
range verification

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals