20-4 Routine Maintenance
DISPLAY PATTERNS test
The display test lets you verify that each pixel and annunciator in the vacuum uorescent
display is working properly. Perform the following steps to run the display test:
1. Display the MAIN MENU by pressing the MENU key.
2. Select TEST, and press ENTER to display the SELF-TEST MENU.
3. Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4. Select DISPLAY-PATTERNS, and press ENTER to start the display test. There are ve
parts to the display test. Each time a front panel key (except EXIT) is pressed, the next
part of the test sequence is selected. The ve parts of the test sequence are as follows:
Checkerboard pattern (alternate pixels on) and all annunciators.
Checkerboard pattern and the annunciators that are on during normal operation.
Horizontal lines (pixels) of the rst digit are sequenced.
Vertical lines (pixels) of the rst digit are sequenced.
Each digit (and adjacent annunciator) is sequenced. All the pixels of the selected
digit are on.
5. When nished, abort the display test by pressing EXIT. The instrument returns to the
FRONT PANEL TESTS MENU. Continue pressing EXIT to back out of the menu
structure.
CHAR SET test
The character set test lets you display all characters. Perform the following steps to run the
character set test:
1. Display the MAIN MENU by pressing the MENU key.
2. Select TEST, and press ENTER to display the SELF-TEST MENU.
3. Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4. Select CHAR-SET, and press ENTER to start the character set test. Press any key
except EXIT to cycle through all displayable characters.
5. When nished, abort the character set test by pressing EXIT. The instrument returns to
the FRONT PANEL TESTS MENU. Continue pressing EXIT to back out of the menu
structure.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals