A-10 Specifications
Case II: Auto-Zero enabled and measuring two functions
Trigger Latency: 225µsec max
Source Configuration: 50µsec max
A/D Conversion: [NPLC Setting
×
(1 / power line frequency)] + 185µsec
Firmware Overhead: 2.3msec for Source V
2.65msec for Source I
Source On Time
Source Configuration + Source Delay + (4
×
A/D Conversion) + Firm-
ware Overhead
Example: Source Delay = 0µsec
NPLC Setting = 0.06 PLC
Power Line Frequency = 60Hz
Source On Time
50µsec + 0 + [(4
×
0.06
×
1/60) + 185µsec] + 2.6msec
7.1msec for Source V
7.45msec for Source I
Case III: Auto-Zero disabled and measuring one function
Trigger Latency: 225µsec max
Source Configuration: 50µsec max
A/D Conversion: [NPLC Setting
×
(1 / power line frequency)] + 185µsec
Firmware Overhead: 300µsec for Source V
640µsec for Source I
Trigger
Latency
Trigger
Event
Source
On
Source On Time
Source
Off
Trigger
Delay
Source
Configuration
Source
Delay
A/D
Conversion
(voltage signal
phase)
A/D
Conversion
(current signal
phase)
A/D
Conversion
(ref phase)
A/D
Conversion
(ref zero phase)
F
i
g
u
r
e A-2
Case II
timing
diagram
Trigger
Latency
Trigger
Event
Source
On
Source On Time
Source
Off
Trigger
Delay
Source
Configuration
Source
Delay
A/D
Conversion
(current signal
phase)
F
i
g
u
r
e A-3
Case III timing
diagram

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals