6-2 Range, Digits, Speed, and Filters
Range and digits
Range
The selected measurement range affects the accuracy of the measurements as well as the
maximum signal that can be measured. Note that with the output off, dashed lines are displayed
(i.e., --.---- µA), to indicate that measurements are not being performed.
Available ranges
The SourceMeter consists of a mainframe and a Remote PreAmp. The mainframe can be
used with or without the Remote PreAmp. However, when not using the Remote PreAmp, the
lower current ranges and higher resistance ranges are not available. Table 6-1 lists the available
ranges for the SourceMeter.
Table 6-1
Model 6430 ranges
Voltage Ranges Current Ranges Ohms Ranges
With Remote
PreAmp
Mainframe
Only
With Remote
PreAmp
Mainframe
Only
With Remote
PreAmp
Mainframe
Only
200V
20V
2V
200mV
200V
20V
2V
200mV
100mA
10mA
1mA
100µA
10µA
1µA
100nA
10nA
1nA
100pA
10pA
1pA
100mA
10mA
1mA
100µA
10µA
1µA
<2Ω*
20Ω
200Ω
2kΩ
20kΩ
200kΩ
2MΩ
20MΩ
200MΩ
2GΩ
20GΩ
200GΩ
2TΩ
20TΩ
<2Ω*
20Ω
200Ω
2kΩ
20kΩ
200kΩ
2MΩ
20MΩ
*Manual ohms mode only