SCPI Command Reference 17-39
:MODE <name>
:CALCulate2:CLIMits:MODE <name> Control Digital I/O port pass/fail output
Parameters <name> = GRADing Output graded pass/fail pattern
SORTing Output sorted pass/fail pattern
Query :MODE? Query Digital I/O pass/fail mode
Description This command controls how limit calculations drive the Digital I/O lines. In
GRADing mode, a reading passes if it is within all of the hi/low limit toler-
ances enabled, assuming that it has passed LIMIT 1 compliance test rst.
The Digital I/O lines will be driven with the rst pattern of the rst compli-
ance, hi, or low failure. Otherwise, the CALC2:CLIM:PASS:SOUR2 pat-
tern will be output.
In SORTing mode, a reading will fail if it fails the compliance test, or is not
within any of the Digital I/O Bands. If the tests pass and only LIMIT 1 is
enabled, the CALC2:CLIM:PASS:SOUR2 pattern will be output. Other-
wise, the rst limit test band that passes will output its LOW:SOUR2 pattern
(UPP:SOUR2 patterns will be ignored). If LIMIT1 fails, its SOUR2 pattern
will be output. If no LIMIT2, 3, 5-12 limit passes, the
CALC2:CLIM:FAIL:SOUR2 pattern will be output.
Clear test results
[:IMMediate]
:CALCulate2:CLIMits:CLEar[:IMMediate] Clears test results and resets Digital I/O
Port
Description This command clears the test results (pass or fail) of the limit tests and
resets the output lines of the Digital I/O port back to the :SOURce2:TTL
settings. See SOURce2.
:AUTO <b>
:CALCulate2:CLIMits:CLEar:AUTO <b> Control auto-clear for test results
Parameters <b> = 1 or ON Enable auto-clear
0 or OFF Disable auto-clear
Query :AUTO? Query state of auto-clear
Description With auto-clear enabled, test results will clear and the output lines of the
Digital I/O port will reset when the :INITiate command is sent to start a new
test sequence.
When disabled, you must use :IMMediate to perform the clear actions.

Table of Contents

Other manuals for Keithley 6430

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
TypeSub-Femtoamp Remote SourceMeter
Compliance Voltage200 V
Maximum Output Voltage200 V
Maximum Output Current100 mA
Compliance LimitProgrammable
InterfaceGPIB, RS-232
Input Impedance>10^14 Ω

Summary

Safety Precautions

3 Basic Source-Measure Operation

Basic Source-Measure Procedure

Details setup for source-measure operations: function, values, limits, and output control.

5 Source-Measure Concepts

Compliance limit

Discusses compliance limits, values, and determination.

Operating boundaries

Covers voltage/current boundaries for source, sink, I-source, V-source, and source-measure modes.

6 Range, Digits, Speed, and Filters

Range and digits

Discusses available ranges, limits, manual/autoranging, and display resolution.

Filters

Provides information on 3-stage filtering to reduce reading noise.

7 Relative and Math

Math Operations

Details math (FCTN) operations: power, offset ohms, varistor, etc.

9 Sweep Operation

Configuring and Running a Sweep

Discusses setup, configuration, delay setting, and performing sweeps.

10 Triggering

Trigger Model

Discusses trigger model, layers, event detection, delay, and device action.

Configuring Triggering

Details how to configure various triggering aspects.

Remote Triggering

Details remote trigger model, commands, and basic triggering example.

11 Limit Testing

Types of Limits

Discusses three limit types: compliance, coarse, fine; summarizes grading and sorting modes.

Configuring and Performing Limit Tests

Describes configuration for limit tests and typical test procedure.

12 Digital I/O Port, Interlock, and Output Configuration

Safety Interlock

Describes using Digital I/O Port as a safety interlock.

14 Status Structure

Status Byte and Service Request (SRQ)

Explains programming Status Byte for SRQs and using serial poll to detect SRQs.

15 Common Commands

Command Summary

Lists IEEE-488.2 common commands used by the SourceMeter.

Command Reference

Provides detailed reference for common commands, excluding status structure.

16 SCPI Signal-Oriented Measurement Commands

Configuring Measurement Function

Provides detailed info on commands to configure measurement function.

Acquiring Readings

Describes commands to acquire post-processed readings, trigger, acquire, and single measurement.

17 SCPI Command Reference

18 Performance Verification

Performing the Verification Test Procedures

Details restoring factory defaults, setting ranges, and output values.

Mainframe Verification

Covers procedures to verify mainframe accuracy alone.

19 Calibration

Mainframe Calibration

Includes procedures to calibrate the mainframe without the Remote PreAmp.

20 Routine Maintenance

Line Fuse Replacement

Covers procedure and part numbers for replacing the line fuse.

F Measurement Considerations

Floating measurement safety concerns

Discusses safety concerns for measurements up to 42V above chassis ground.

Leakage currents and guarding

Explains reducing leakage currents using insulators, humidity control, and guarding.