SCPI Command Reference 17-89
:SYSTem subsystem
The SYSTem subsystem contains miscellaneous commands that are summarized in
Table 17-8.
Default conditions
:PRESet
:SYSTem:PRESet Return to :SYSTem:PRESet defaults
Description This command returns the instrument to states optimized for front panel
operation. :SYSTem:PRESet defaults are listed in the SCPI tables
(Tables 17-1 through 17-10).
:POSetup
:SYSTem:POSetup <name> Program power-on defaults
Parameters <name> = RST Power-up to *RST defaults
PRESet Power-up to :SYSTem:PRESet defaults
SAV0 Power-up to setup stored at memory location 0
SAV1 Power-up to setup stored at memory location 1
SAV2 Power-up to setup stored at memory location 2
SAV3 Power-up to setup stored at memory location 3
SAV4 Power-up to setup stored at memory location 4
Query :POSetup? Query power-on setup
Description This command is used to select the power-on defaults. With RST selected,
the instrument powers up to the *RST default conditions. With PRES
selected, the instrument powers up to the :SYStem:PRESet default condi-
tions. Default conditions are listed in the SCPI tables (Tables 17-1 through
17-10).
With the SAV0-4 parameters specied, the instrument powers-on to the
setup that is saved in the specied location using the *SAV command.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals