SCPI Command Reference 17-67
:AUTO <b>
:SOURce[1]:CURRent:RANGe:AUTO <b> Select auto range for I-Source
:SOURce[1]:VOLTage:RANGe:AUTO <b> Select auto range for V-Source
Parameters <b> = 0 or OFF Disable auto range
1 or ON Enable auto range
Query AUTO? Query state of auto range
Description This command is used to enable or disable auto range for the specified
source. When enabled, the instrument will automatically select the most
sensitive range for the specified source level. When disabled, the instrument
will use the range that the instrument is currently on.
Auto range will be disabled if a fixed range is selected. See the previous
command.
Both *RST and :SYSTem:PREset enables source auto range. When the
SourceMeter goes into the local state, source auto range disables.
Set amplitude for fixed source
[:IMMediate][:AMPLitude] <n>
:SOURce[1]:CURRent[:LEVel][:IMMediate][:AMPLitude] <n> Set fixed I-Source amplitude immediately
:SOURce[1]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <n> Set fixed V-Source amplitude immediately
Parameters <n> = -105e-3 to 105e-3 Set I-Source amplitude (amps)
-210 to 210 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -105e-3A or -210V
MAXimum +105e-3A or +210V
Query :CURRent? Query programmed amplitude of I-Source
:CURRent? DEFault Query *RST default amplitude
:CURRent? MINimum Query lowest allowable amplitude
:CURRent? MAXimum Query highest allowable amplitude
:VOLTage? Query programmed amplitude of V-Source
:VOLTage? DEFault Query *RST default amplitude
:VOLTage? MINimum Query lowest allowable amplitude
:VOLTage? MAXimum Query highest allowable amplitude
Description This command is used to immediately update the amplitude of a fixed
source. This command is not valid if the list or sweep mode is selected.
NOTE The sourcing :MODE command is used to select a fixed source. See “Select sourcing
mode.”