17-70 SCPI Command Reference
Set delay
:DELay <n>
:SOURce[1]:DELay <n> Manually set source delay 
Parameters <n> = 0 to 999.9999 Specify delay in seconds
MINimum 0 seconds
MAXimum 999.9999 seconds
DEFault 0.003
Query :DELay? Query delay
:DELay? DEFault Query *RST default delay
:DELay? MINimum Query lowest allowable delay
:DELay? MAXimum  Query highest allowable delay
Description This command is used to manually set a delay (settling time) for the source. 
After the programmed source is turned on, this delay occurs to allow the 
source level to settle before a measurement is taken. Note that this delay is 
the same for both the I-Source and V-Source. 
Do not confuse this source delay with the trigger delay. The source delay is 
part of the device action (SDM cycle) while the trigger delay occurs before 
the device action. See Section 10, Trigger model for more information. 
Auto delay can instead be used to automatically set the source delay. See the 
next command.
:AUTO <b>
:SOURce[1]:DELay:AUTO <b> Enable/disable auto delay for source 
Parameters <b> = 0 or OFF Disable auto delay
1 or ON Enable auto delay
Query :AUTO? Query state of auto delay
Description This command is used to enable or disable auto delay. When enabled, the 
instrument will automatically select a delay period that is appropriate for the 
present source/measure setup configuration. See Table 3-2. *RST and 
SYST:PRES default is OFF.