Performance Verification 18-31
2. Congure the Model 6430 ohms function for the 2-wire sense and guard modes as
follows:
Press CONFIG then MEAS . The instrument will display the following:
CONFIG OHMS
SOURCE GUARD
Select GUARD, and then press ENTER. The following will be displayed:
GUARD
OHMS CABLE
Select CABLE, and then press ENTER.
Press EXIT to return to normal display.
WARNING Hazardous voltage may be present on SHIELD when using guarding.
3. Press MEAS to select the ohms measurement function, and make sure the source out-
put is turned on.
4. Verify ohms measurement accuracy for each of the resistance values listed in
Table 18-16. For each measurement:
Connect the correct standard resistance directly to the IN/OUT HIGH jack.
Select the appropriate ohms measurement range with the RANGE keys.
Verify that the Model 6430 resistance reading is within the limits given in the
table.
Table 18-16
Remote PreAmp 2T
and 20T
range measurement accuracy limits
Model 6430 range Standard resistance
1
Model 6430 resistance reading limits
2
(1 year, 18°-28°C)
2T
20T
1T
10T
0.99168 to 1.00832T
9.79300 to 10.2070
1
Nominal values shown.
2
Calculated using Model 6430 normal accuracy specications and nominal resistance values.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals