SCPI Command Reference 17-107
OUTPut <event list>
:ARM[:SEQuence[1]][LAYer[1]][:TCONgure]:OUTPut <event list> Arm layer events
:TRIGger[:SEQuence[1]][:TCONgure]:OUTPut <event list> Trigger layer events
Parameters Arm Layer Triggers
<event list >: TENTer Trigger on entering trigger layer
TEXit Trigger on exiting trigger layer
NONE Disable arm layer output trigger
Trigger Layer Triggers
<event list>: SOURce Output trigger after source level is set
DELay Output trigger after delay period
SENSe Output Trigger after measurement
NONE Disable trigger layer triggers
NOTE Each event in the list must be separated by a comma (i.e., :arm:output source, delay,
sense).
Query :OUTPut? Query output trigger event(s)
Description This command is used to specify when trigger pulses occur on the specied
output trigger line of the Trigger Link (:OLINe).
Arm Layer Triggers With TEXit selected, an output trigger will occur
when exiting the trigger layer. With TENTer selected, an output trigger will
occur when entering the trigger layer. With NONE selected, the arm layer
output trigger is disabled.
Trigger Layer Triggers You can specify from one to all three events. Each
event in the list must be separated by a comma (,).
The SOURce, DELay and MEASure events refer to the Source-Delay-
Measure (SDM) cycle. This is the Device Action in the Trigger Model. See
Figures 10-7 and 10-8. With SOURce specied, an output trigger occurs
after the source is set. With DELay specied, an output trigger occurs after
the delay period. With MEASure specied, an output trigger occurs after the
measurement.
NOTE When disabling triggers, the NONE parameter must be sent alone (i.e., trig:outp
none). If it is listed with any of the other event parameters, NONE will be ignored.

Table of Contents

Other manuals for Keithley 6430

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
TypeSub-Femtoamp Remote SourceMeter
Compliance Voltage200 V
Maximum Output Voltage200 V
Maximum Output Current100 mA
Compliance LimitProgrammable
InterfaceGPIB, RS-232
Input Impedance>10^14 Ω

Summary

Safety Precautions

3 Basic Source-Measure Operation

Basic Source-Measure Procedure

Details setup for source-measure operations: function, values, limits, and output control.

5 Source-Measure Concepts

Compliance limit

Discusses compliance limits, values, and determination.

Operating boundaries

Covers voltage/current boundaries for source, sink, I-source, V-source, and source-measure modes.

6 Range, Digits, Speed, and Filters

Range and digits

Discusses available ranges, limits, manual/autoranging, and display resolution.

Filters

Provides information on 3-stage filtering to reduce reading noise.

7 Relative and Math

Math Operations

Details math (FCTN) operations: power, offset ohms, varistor, etc.

9 Sweep Operation

Configuring and Running a Sweep

Discusses setup, configuration, delay setting, and performing sweeps.

10 Triggering

Trigger Model

Discusses trigger model, layers, event detection, delay, and device action.

Configuring Triggering

Details how to configure various triggering aspects.

Remote Triggering

Details remote trigger model, commands, and basic triggering example.

11 Limit Testing

Types of Limits

Discusses three limit types: compliance, coarse, fine; summarizes grading and sorting modes.

Configuring and Performing Limit Tests

Describes configuration for limit tests and typical test procedure.

12 Digital I/O Port, Interlock, and Output Configuration

Safety Interlock

Describes using Digital I/O Port as a safety interlock.

14 Status Structure

Status Byte and Service Request (SRQ)

Explains programming Status Byte for SRQs and using serial poll to detect SRQs.

15 Common Commands

Command Summary

Lists IEEE-488.2 common commands used by the SourceMeter.

Command Reference

Provides detailed reference for common commands, excluding status structure.

16 SCPI Signal-Oriented Measurement Commands

Configuring Measurement Function

Provides detailed info on commands to configure measurement function.

Acquiring Readings

Describes commands to acquire post-processed readings, trigger, acquire, and single measurement.

17 SCPI Command Reference

18 Performance Verification

Performing the Verification Test Procedures

Details restoring factory defaults, setting ranges, and output values.

Mainframe Verification

Covers procedures to verify mainframe accuracy alone.

19 Calibration

Mainframe Calibration

Includes procedures to calibrate the mainframe without the Remote PreAmp.

20 Routine Maintenance

Line Fuse Replacement

Covers procedure and part numbers for replacing the line fuse.

F Measurement Considerations

Floating measurement safety concerns

Discusses safety concerns for measurements up to 42V above chassis ground.

Leakage currents and guarding

Explains reducing leakage currents using insulators, humidity control, and guarding.