2400 Series SourceMeter
®
User’s Manual Source-Measure Concepts 6-27
Figure 6-12
V-Source output characteristics
210V21V
105mA
1.05A
A. Model 2400
1100V21V
21mA
1.05A
B. Model 2410
63V21V
1.05A
3.15A
C. Model 2420
21V
1.05A
3.15A
D. Models 2425 and 2430
10.5A
Model 2430
Pulse Mode
DC Mode
105V
Limit I Limit I
Limit I
Limit I
Source V
Source V
Source VSource V
42V10.5V
1.05A
5.25A
E. Model 2440
Limit I
Source V