12-24 Limit Testing 2400 Series SourceMeter
®
User’s Manual
Limit test programming example
Diode breakdown voltage test is an example that readily lends itself to pass/fail analysis.
This test verifies the reverse and often the forward voltage at which the device begins to
show a large deviation in current for a small deviation in voltage. The test is performed by
sourcing a specified current level and then measuring the resulting voltage drop. The volt-
age drop is then compared with one set of limits to determine if the diode passes or fails
and should be discarded. Voltage measurements for failing diodes are also compared
against a more restricted range of limits to determine if they should be routed to QA
(Quality Assurance) for further analysis.
Command* Description*
:CALCulate2:CLIMitis:FAIL:SMLocation <location> Specify fail source memory location. (Location =
NRf (memory#) or NEXT (next location).
:CALCulate2:CLIMits:BCONtrol <name> Control I/O port pass/fail update. Name =
IMMediate (at failure) or END (end of sweep).
:CALCulate2:CLIMits:CLEar Clear test results, reset I/O port.
:CALCulate2:CLIMits:CLEar:AUTO <state> Enable/disable auto-clear (state = ON or OFF).
:CALCulate2:CLIMits:MODE <name> Select Digital I/O control mode (name =
GRADing or SORTing).
* LIMitX = LIMit2, LIMit3, LIMit5 through LIMit12. LIMit4 (Contact check option) not shown (see Table 18-1 and Appendix F).
Table 12-2
Commands to control Digital I/O port for limit testing
Command Description
:SOURce2:BSIZe <n>
:SOURce2:TTL <NRf> | <NDN>
:SOURce2:TTL4:MODE <name>
:SOURce2:TTL4:BSTate <state>
:SOURce2:CLEar
:SOURce2:CLEar:AUTO <state>
:SOURce2:CLEar:AUTO:DELay <n>
Set Digital I/O port bit size (n = 3 or 4).
Set I/O port bit pattern (NRf | NDN = pattern).
Set Digital I/O line 4 mode (name = EOTest or BUSY).
Set BUSY and EOT polarity (HI or LO).
Clear digital output lines.
Enable/disable I/O port auto clear (state = ON or OFF).
Set auto-clear delay (n = delay).
Table 12-1 (continued)
Limit commands
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2410 and is the answer not in the manual?

Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals